线路泄漏测试(电气安全测试标准要求的第四项测试)

发布时间:2022-02-13 点击数:472

技术支持详情

The Line Leakage Test is the fourth test required by the electrical safety testing standards.

The Line Leakage Test is done through a milliammeter with a 1750Ω impedance and a time constant of 225µs (for the EN60335-1 standard) equivalent to the human body. This milliammeter is connected between the neutral conductor and the ground of the DUT for the first line leakage test and then between the phase conductor and the ground of the DUT for the second line leakage test.

A- Line leakage test - Operating mode without insulation transformer:
If the DUT is connected to the earth, the differential breaker of the electrical installation of the building will operate (230 Volts on 1750Ω will drive approximately 130mA).

If the DUT is not connected to the earth, the differential breaker will not operate, but when doing the second measurement (phase/ground), the voltage difference between the housing of the DUT and the earth will be 230 Volts : this voltage is dangerous for the operator. When doing the first measurement, (neutral/ground), the voltage difference between the housing of the DUT and the earth will be only a few volts : this is less dangerous, but not recommended. During this operation mode, it is advised to deny the access to the DUT to anybody without ability, by the use of a safety enclosure or a mechanical, respectively optical barrier.


B- Line leakage test - Operating mode with insulation transformer:

If the DUT is connected to the earth, the differential breaker of the electrical installation of the building won't operate, because of the galvanis insulation between the mains and the DUT. In this operating mode, the ground of the DUT can be permanently connected to the earth, which is safer for the operator.


clip_image002_002

泄漏电流测试


测试目的:检测设备通电情况下,可接触部件与电源之间的泄漏电流值

测试方法:设备通过隔离变压器供电,检测此时设备相线,中线和地线之间的泄漏电流值是否满足标准要求

推荐型号:FMG500  FXS501


测试标准

EN61010-1

EN60335-1

EN60950

EN60598-1

EN60601

EN61131-2

电压

 ________

U nominal

Un x 1.06

U nominal

(<1kHz)

U nominal

 Un x 1,1

  _________

电流 阈值

  ________

0.5mA

3.5mA

0.5mA

3.5mA

0.5–1 mA

5mA max

10µA to10mA

  _________

测试电路

  ________

Figure 4

Figure 4

Figure 4 & 5

   Figure EN60601

  _________